Human Hematology Analyzer
                                        - PE-6100
- PE-7100
3-Part Fully Auto Hematology Analyzer – PE-6100
   
                                                                                    - Automated & User-friendly Interface
- Accurate Results & Low Sample Consumption
- Easy Maintenance & Environment-friendly System
- Perfect QC System

| Principle | Electrical resistance for counting, cyanide free method for hemoglobin | 
| Sample Consumption | Venous mode: 9.6μL, Capillary mode: 9.6μL, Pre-diluted mode: 20μL | 
| Working Mode | Double passages counting detection + independent HGB measurement system; 24 hours with auto standby | 
| Operation Mode | USB keyboard & mouse (PE-6000Plus: Touch screen) | 
| Throughput | ≥60 samples/hour | 
| Storage | 50,000 sample results inclusive of the histograms | 
| QC Mode | L-J QC mode | 
| Reference Value | Baby, Kid, Woman, Man, Normal and Customization | 
| Cross Contamination Ratio | WBC/RBC/HGB﹤0.5%, PLT﹤1% | 
| Interface | 4×USB, 1×ethernet, 1×VGA (for external monitor) | 
| Printer | Built-in thermal recorder, optional external printer, eight report formats | 
| Environment Requirement | Temperature: 15℃~35℃; Humidity: 10%-90% | 
| Parameters | 3-part differential of WBC; 20 parameters (WBC, LYM#, MID#, GRAN#, LYM%, MID%, GRAN%, RBC, HGB, HCT, MCV, MCH, MCHC, RDW-SD, RDW-CV, PLT, MPV, PDW, PCT, P-LCR) + 3 color histograms (WBC/RBC/PLT) | 
5-Part Fully Auto Hematology Analyzer – PE-7100
   
                                                                                    
                                                                                    - Electrical Impedance method for determining the RBC and PLT data
- Colorimetric method for determining the HGB
- Laser-based flow cytometry for determining the WBC data

| Principle | 23 parameters+8 Research parameters +4 scatter diagrams and 3 histograms: WBC/RBC/PLT; 5 Diff of WBC | 
| Sample Volume | Venous Mode/Capillary Mode/Pre-diluted Mode≤20μL | 
| Operation Mode | Touch screen operation, also support USB keyboard & mouse; Based on Linux system | 
| Throughput | 60 samples/h, 24 hours work with auto standby | 
| Storage | 50,000 sample results inclusive of the histograms | 
| Application | Baby/kid/women/man/normal/customized | 
| Quality control | L-J; X-B | 
| Display | 10.4'' touch screen | 
| Language | Multi-Language supported, user can on-line switch interface language | 
| Calibration mode | Manual calibration and auto calibration | 
| Weight | 26.5KG | 
| Dimension | 364×498×431(mm) | 
| Working environment | Temperature: 10-30℃, Humidity≤70% | 
| Test mode | CBC+DIFF, CBC | 